General Specifications

Operating environment

Indoor use, pollution level 2, altitude up to 2000 m

Operating temperature and humidity range

−20°C to 50°C (−4°F to 122°F), 80% RH or less (non-condensing)

When using below 0°C (32°F), pre-warm the device in an environment between 0°C and 50°C (32°F to 122°F) beforehand.

Storage temperature and humidity range

−20°C to 50°C (−4°F to 122°F), 80% RH or less (non-condensing)

Conforming standards

Safety: EN 61010

EMC: EN 61326 Class A

Complying standards

(PW9006 option)

The instrument in IEC measurement mode complies with IEC 61000-4-7:2002 (Ed. 2.0).

The instrument in IEC measurement mode complies with IEC 61000-4-15:2010 (Ed. 2.0).

Vibration resistance

JIS D 1601:1995 5.3 (1) Type 1: Passenger cars, Condition: Equivalent to Type A

Vibration acceleration: 45 m/s2 (4.6 G) for 4 h in the X direction and 2 h in the Y and Z directions

Power supply

  • Commercial power supply

Rated supply voltage: 100 to 240 V AC

(Assuming voltage fluctuation of ±10%)

Rated power-supply frequency: 50 Hz, 60 Hz

Anticipated transient overvoltage: 2500 V

Maximum rated power: 120 VA

Typical power consumption (reference value): 47 W

(Conditions: Power supply voltage of 100 V/60 Hz. Voltage of 800 V DC and current of 200 A DC (CT6834) measured on all channels.)

  • DC power supply (optional)

Rated supply voltage:

10.5 V to 28 V DC (Operating temperature range: −20°C to 40°C (−4°F to 104°F))

10.5 V to 20 V DC (Operating temperature range: 40°C to 50°C (104°F to 122°F))

Maximum rated power: 95 VA

Backupbattery life

Lithium battery: About 10 years (Reference value at 23°C (73°F))

The battery is used to back up the time and setting conditions.

Dimensions

Approx. 361W × 176H × 135D mm (14.2"W × 6.9"H × 5.3"D) (excluding protruding parts)

Weight

Approx. 4.6 kg (10.1 lb.) (for PW4001-05)

Product warranty duration

3 years

Accuracy guarantee conditions

Accuracy guarantee duration: 12 months

(The accuracy guarantee duration for voltage, current, and power measurements, as well as for voltage accuracy of the motor analysis option, is 6 months. The 12-month accuracy is calculated by multiplying the reading error specified in each accuracy specification by 1.5.)

Accuracy guarantee temperature and humidity range: 23°C ±3°C (73°F ±5°F), 80% RH or less

Warm-up time: 30 minutes or longer

Other conditions: Within the effective measurement ranges, sine waveforms or DC input, a line-to-earth voltage of 0 V, after zero adjustment has been performed and a change in ambient temperature does not exceed ±1°C after zero adjustment