Auto-hold conditions

When both the following conditions are satisfied, the instrument freezes the readout.

  • The measured value does not fluctuate beyond the variation range for a certain period.

  • The measured value exceeds the threshold value.

The instrument continues to freeze the readout until the auto-hold conditions are satisfied again.

 

  1. (1)

    The instrument does not freeze the readout automatically. (The measured value does not become less than the threshold value.)

  2. (2)

    The instrument does not freeze the readout automatically. (The measured value does not exceed the threshold value.)

 

The variation range and threshold value may vary depending on the range.

Range

Variation width

Threshold value

60.00 mA

400 counts or less

150 counts

600.0 mA

6.000 A

60.00 A

500 counts or less

600.0 A